Table 1 EDS results of the three cases of early implant failures

From: Analysis of acute sinusitis-related early failed implant surface: a combined histological, electron microscopy, and X-ray spectroscopy approach

Implant no.

EDS-examined position

EDS result (C: weight percentage)

Surface morphology (SEM results)

C ≥ 10%

10% > C ≥ 1%

C < 1%

01

Bone tissue in the upper region (0101-B)

C: 38.43%

O: 32.90%

N: 16.08%

Ca: 7.87%

Tl: 2.42%

Si: 1.90%

S: 0.41%

- Bone tissue showed irregular structure with no presence of cells or bone lacunae.

- Low level of Ca and no P signal.

- High levels of C, N, and O → large portion of organic content.

- Noticeable levels of Tl and Si.

- Trace S signal.

Implant surface at the first thread of the fixture (0102-U)

Ti: 56.31%

O: 21.03%

C: 16.90%

Ca: 4.16%

Si: 1.60%

 

- SEM image showed an irregular morphology of a sandblasted and acid-etched surface, with signs of oxidation.

- Titanium surface with a high level of O → oxidized surface.

- Minor level of Ca → low osseointegration on the implant surface.

- Si signal.

Graft material on the upper region (0103-G)

O: 34.26%

C: 23.08%

Ca: 23.03%

N: 17.72%

 

Na: 0.97%

S: 0.94%

- SEM image showed an irregular bone structure with suggested grafting material particles.

- There was no detection of the P signal.

- High levels of N and O suggested that this is calcified organic material.

- Trace Na and S signals.

02

Implant surface in the upper region (0204-U)

C: 35.79%

Au: 26.73%

O: 15.52%

Ti: 8.98%

Si: 6.48%

Na: 3.27%

Cl: 3.24%

 

- Heterogeneous surface with organic particles.

- Noticeable high level of Au.

- Low level of Ti but high levels of C and O → large portion of organic matter.

- Si, Na, and Cl were detected to a minor degree.

Implant surface in the middle region (0205-M)

Ti: 51.85%

O: 15.14%

Zr: 11.10%

C: 8.71%

Ca: 8.12%

Au: 3.03%

Na: 1.60%

Si: 0.46%

- Heterogeneous surface with micro-fissures.

- Presence of a thin bone layer and organic matter.

- Titanium surface with a high level of O → oxidized surface.

- Noticeably high level of Zr.

- Detection of Au and Na at a low level.

- Trace amount of Si.

Implant surface in the apical region (0206-A)

Ti: 71.72%

O: 9.88%

Au: 7.99%

C: 7.75%

Si: 1.27%

Na: 1.02%

Ca: 0.36%

- While the EDS result showed the rich Ti content, the SEM image of the surface revealed an irregular morphology with fibers and micro-debris.

- Rich of Ti.

- Au was detected to a noticeable degree.

- Minor signs of Si and Na.

- Trace amount of Ca.

03

Implant surface in the upper region (0307-U)

Au: 40.11%

C: 31.69%

Si: 11.86%

O: 10.14%

Ti: 4.26%

Na: 1.94%

 

- The fixture’s surface was covered with fibers and heterogeneous organic matter.

- Significantly high level of Au.

- High levels of C and O → large portion of organic matter.

- Minor Na signal.