Skip to main content

Table 1 Accuracy of printed implants and surgical bone resection guides

From: Investigating the accuracy of mandibulectomy and reconstructive surgery using 3D customized implants and surgical guides in a rabbit model

 

Pre- and postoperative implants (mm)

Pre- and postoperative surgical bone resection guides (mm)

Planned

Output

Planned

Output

1

10.00

10.01

10.10

10.05

2

10.00

10.02

10.10

10.08

3

10.00

10.04

10.10

10.12

4

10.00

10.02

10.10

10.07

5

10.00

10.03

10.10

10.09

6

10.00

10.02

10.10

10.09

7

10.00

10.00

10.10

10.09

8

10.00

10.01

10.10

10.11

9

10.00

10.00

10.10

10.19

10

10.00

10.04

10.10

10.19

11

10.00

10.00

10.10

10.11

12

10.00

10.05

10.10

10.08

13

10.00

10.05

10.10

10.11

14

10.00

10.02

10.10

10.14

15

10.00

10.02

10.10

10.14

16

10.00

10.00

10.10

10.15

17

10.00

10.01

10.10

10.05

18

10.00

10.02

10.10

10.06

19

10.00

10.02

10.10

10.12

20

10.00

10.01

10.10

10.12

Average

10.00

10.02

10.10

10.11

SD

0.00

0.02

0.00

0.04