Skip to main content

Table 2 Accuracy of printed implants and surgical bone resection guides: equivalence comparison

From: Investigating the accuracy of mandibulectomy and reconstructive surgery using 3D customized implants and surgical guides in a rabbit model

Variable

Mean

SE

95%CIs

P value

(Equi. interval: − 0.01–0.01)

P value

(Equi. interval: − 0.02–0.02)

P value

(Equi. interval: − 0.03–0.03)

Pre- and postoperative implant: planned

10

0

10–10

   

Pre- and postoperative implant: output

10.019

0.004

10.013–10.026

   

Difference between pre- and postoperative findings: output–planned

0.019

0.004

0.013–0.026

0.993

0.444

0.004

Pre- and postoperative surgical bone resection guides: planned

10.1

0

10.1–10.1

   

Pre- and postoperative surgical bone resection guides: output

10.108

0.009

10.09–10.126

   

Difference in pre- and postoperative surgical bone resection guides: output−planned

0.008

0.009

− 0.01–0.026

0.413

0.099

0.012

  1. CI confidence interval, SE standard error